EasyManuals Logo

ADVANTEST R3131 User Manual

Default Icon
261 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #221 background imageLoading...
Page #221 background image
R3131 Spectrum Analyzer Operation Manual
4.1 GPIB Remote Programming
4-34 Aug 28/98
(1) Sample Programs for PC9801 series (GPIB address = 8)
Example PC-5: Output marker level (numerical variable)
Sample result: MARKER LEVEL=-16.22
Example PC-6: Output center frequency (character variable)
Sample result: 30.000E+6
Example PC-7: Output level and its unit
Sample result: 0.0E+0:0
Example PC-8: Execute 6 dB down and then output frequency and level values (multiple items).
Sample result: MARKER FREQ=400000: MARKER LEVEL=1.16
10 ISET IFC:ISET REN
20 PRINT @8;”CF30MZ SP1MZ MK30MZ” Center frequency, frequency span, marker
ON
30 PRINT @8;”ML?” Marker level?
40 PRINT @8;ML Read marker level.
50 PRINT ”MARKER LEVEL=”,ML Display the result.
60 STOP
70 END
10 ISET IFC:ISET REN
20 PRINT @8;”CF?”
30 INPUT @8;CF$ Read center frequency.
40 PRINT CF$ Display the result.
50 STOP
60 END
10 ISET IFC:ISET REN
20 PRINT @8;”RL?”
30 INPUT @8;RE$ Read REF level.
40 PRINT @8;”UN?”
50 INPUT @8;UN Read unit for level
60 PRINT RE$,”:”,UN Display the result
70 STOP
80 END
10 ISET IFC:ISET REN
20 PRINT @8;”CF30MZ SP20MZ” Set center frequency and frequency span.
30 PRINT @8;”MKBW6DB PS XDB” Execute 6 dB down.
40 PRINT @8;”MFL?” Read frequency and level for marker position
at the same time.
50 INPUT @8;MF,ML
60 PRINT ”MARKER FREQ=”;MF;”:MARKER LEVEL=”;ML
70 STOP
80 END

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3131 and is the answer not in the manual?

ADVANTEST R3131 Specifications

General IconGeneral
BrandADVANTEST
ModelR3131
CategoryMeasuring Instruments
LanguageEnglish