REMOTE OPERATION CARRIER FREQUENCY COMMANDS
4-38
[SOURce]:FREQuency:SWEep:MANual
Description: Sets a new carrier frequency whilst a sweep is paused.
Parameters: <numeric_value>
Valid values: <NRf>(Hz) | MAXimum | MINimum | UP | DOWN
Set by value, to maximum or minimum, or stepping up or down.
This command is available only when FREQ:MODE SWEep is selected, and sweep
operation is not in progress (PAUSED or WAITING FOR TRIGGER). The frequency
value should be limited to the range determined by FREQ:SWEep:STARt and
FREQ:SWEep:STOP.
[SOURce]:FREQuency:SWEep:MANual?
Description: Queries the carrier frequency set during a paused sweep.
Parameters: None
Response: <NR2>
Returned values: Carrier frequency in Hz
[SOURce]:FREQuency:SWEep:SPACing
Description: Sets the carrier sweep step points to either linear or logarithmic spacing.
Parameters: <CPD>
Valid values: LINear | LOGarithmic
*RST sets: LIN
[SOURce]:FREQuency:SWEep:SPACing?
Description: Queries whether carrier sweep step points have linear or logarithmic spacing.
Parameters: None
Response: <CRD>
Returned values: LIN | LOG