REMOTE OPERATION POWER COMMANDS
4-152
[SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude]:STEP[:INCRement]
Description: Sets the step size for carrier level.
Parameters: <numeric_value>
Valid values: <NRf>(dB) | MAXimum | MINimum
*RST sets: 1 dB
[SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude]:STEP[:INCRement]?
Description: Queries the step size for carrier level.
Parameters: None
Response: <NR2>
Returned values: Carrier level step size in dB
[SOURce]:POWer:LIMit[:AMPLitude]
Description: Sets the maximum RF level limit.
Parameters: <numeric_value>
Valid values: <NRf>(dB) | MAXimum | MINimum
*RST sets: MAX
[SOURce]:POWer:LIMit[:AMPLitude]?
Description: Queries the maximum RF level limit.
Parameters: None
Response: <NR2>
Returned values: Power level limit, in the units set on page 4-172