Agilent 1200 Series Autosampler Reference Manual 153
Repairing the Autosampler 4
3 Remove the metering head assembly (see “Gripper
Arm” on page 128).
4 Remove the four fixing screws, and remove the motor.
5 .Feed the motor cable through to the SUD board. Connect
the cable to the connector, labelled “Needle M”.
6 Secure the motor in place with the four fixing screws.