© Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 10-133
Chapter 10: Site Preparation – Power Requirements for Older Systems: Tester Power Requirements (Series I/II)
327X 317X 307X
3
xDSL/POTS
Bay
Opts/Current
4
Testhead Support Bay
Single-Phase
Non-Earthed
AWX 200 NA 36 42 20 15
0EB 220 NA 36 42 20 15
0EJ 230 NA 36 42 20 15
0EC 240 NA 36 42 20 15
Single-Phase Center-Tap
Neutral
AWY 100 / 200 25 43 42 24 15
0E3 110 / 220 25 43 42 24 15
0EG 115 / 230 25 43 42 24 15
0EH 120 / 240 25 43 42 24 15
1. Calculated with a fully-loaded 3X7X system with 3 STC cards in every module.
2. DUT Power supplies are configured for 200-240 volts ac.
3. 2- and 3- phase 307X systems only require one PDU. Single Phase systems require one PDU for the testhead and one for the support bay.
Table 10-32 3070 Series II family power requirements with E1135A/B PDU (continued)
System
Power Option Voltage Full-Load Amps (FLA)
1
for:
2
L1
L2
G
L1
L2
G
N