© Agilent Technologies 1989–2002 Agilent 3070 / 79000 Site Preparation 4-11
Chapter 4: Site Preparation – Structural Requirements: Moving Access Requirements
Ramp Requirements
When moved on its casters, the testhead will negotiate
ramps with inclines up to 8 percent before the leveling
feet drag on the floor.
Hallway and Door Width Requirements
As a rough guide, if you have room to move a 4-foot by
8-foot (1.25 m x 2.5 m) sheet of plywood, parallel to the
floor, through hallways and doors, you will be able to
move the crated testhead. If you cannot move the
testhead crate all the way, you must remove it from its
crate in a receiving area and push it on its casters to the
destination.