70 OUTPUT 722;"TARM SGL" !TRIGGER READINGS
80 END
High-Speed DCI Example The following program measures DC current at the fastest possible rate.
10 OUTPUT 722;"PRESET FAST" !DCV, 10V RANGE, TARM SYN, TRIG AUTO
20 OUTPUT 722;"APER 1.4E-6"I !LONGEST INTEGRATION TIME POSSIBLE FOR
25 !MAXIMUM READING RATE
30 OUTPUT 722;"MFORMAT SINT" !SINT MEMORY FORMAT
40 OUTPUT 722;"MEM FIFO" !ENABLE READING MEMORY
50 OUTPUT 722;"DCI 100E-3" !DC CURRENT, 100mA RANGE
60 OUTPUT 722;"NRDGS 5000 AUTO" !5000 READINGS/TRIGGER, AUTO SAMPLE EVENT
70 OUTPUT 722;"TARM SGL" !TRIGGER READINGS
80 END
Fast Synchronous
ACV/ACDCV Example
The following program measures AC voltage using the synchronous method
at the fastest possible rate (approximately 10 readings per second). This
program can be adapted to AC+DC voltage by using the ACDCV command
instead of the ACV command in line 50.
10 OUTPUT 722;"PRESET FAST" !TARM SYN, TRIG AUTO
20 OUTPUT 722;"MFORMAT SINT" !SINT MEMORY FORMAT
30 OUTPUT 722;"MEM FIFO" !ENABLE READING MEMORY
40 OUTPUT 722;"SETACV SYNC" !SYNCHRONOUS AC MEASUREMENT METHOD
50 OUTPUT 722;"ACV 10,2" !AC VOLTS, 10V RANGE, 2% RESOLUTION
60 OUTPUT 722;"ACBAND 5E3,8E3" !SIGNAL BETWEEN 5kHz AND 8kHz
70 OUTPUT 722;"NRDGS 20, AUTO" !20 READINGS/TRIGGER, AUTO SAMPLE EVENT
80 OUTPUT 722;"TARM SGL" !TRIGGER READINGS
90 END
Fast Random
ACV/ACDCV Example
The following program measures AC voltage using the random method at
the fastest possible rate (approximately 45 readings per second). This
program can be adapted to AC+DC voltage by using the ACDCV command
instead of the ACV command in line 50.
10 OUTPUT 722;"PRESET FAST"" !TARM SYN, TRIG AUTO
20 OUTPUT 722;"MFORMAT SINT !SINT MEMORY FORMAT
30 OUTPUT 722;"MEM FIFO" !ENABLE READING MEMORY
40 OUTPUT 722;"SETACV RNDM" !RANDOM AC MEASUREMENT METHOD
50 OUTPUT 722;"ACV 10 6" !AC VOLTS, 10V RANGE, 6% RESOLUTION
60 OUTPUT 722;"ACBAND 10E3,20E3” !SIGNAL BETWEEN 10kHz AND 20kHz
70 OUTPUT 722;"NRDGS 100, AUTO” !100 READINGS/TRIGGER, AUTO SAMPLE EVENT
80 OUTPUT 722;"TARM SGL" !TRIGGER READINGS
90 END
Fast Analog ACV/ ACDCV
Example
The following program measures AC voltage using the analog method at a
fast rate. This program uses the default delay time. You can achieve faster
reading rates by specifying a shorter delay time; the resulting settling time,
however, may not produce accurate measurements. You can also achieve
unspecified faster reading rates by specifying less integration time in line 60.
This program can be adapted to AC+DC voltage by using the ACDCV
command instead of the ACV command in line 50.
10 OUTPUT 722;"PRESET FAST"" !TARM SYN, TRIG AUTO
20 OUTPUT 722;"MFORMAT SINT !SINT MEMORY FORMAT
30 OUTPUT 722;"MEM FIFO" !ENABLE READING MEMORY, FIFO MODE
40 OUTPUT 722;"SETACV ANA" !ANALOG AC MEASUREMENT METHOD

Table of Contents

Other manuals for Agilent Technologies 3458A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3458A and is the answer not in the manual?

Agilent Technologies 3458A Specifications

General IconGeneral
BrandAgilent Technologies
Model3458A
CategoryMultimeter
LanguageEnglish

Related product manuals