Appendix A Specifications 297
8 / System Specifications
Function-Range-Measurement
The time required to program via GPIB a new measurement configuration, trigger a reading, and return the
result to a controller with the following instrument setup: PRESET FAST; DELAY 0; AZERO ON;
OFORMAT SINT; INBUF ON; NPLC 0.
Selected Operating Rates
2
Memory
TO - FROM Configuration Description
GPIB Rate
1
Subprogram Rate
DCV £ 10 V to DCV £10 V 180/sec 340/sec
any DCV / OHMS to any DCV / OHMS 85/sec 110/sec
any DCV/OHMS to any DCV/ OHMS with DEFEAT ON 150/sec 270/sec
TO or FROM any DCI 70/sec 90/sec
TO or FROM any ACV or ACI 75/sec 90/sec
Conditions Rate
DCV Autorange Rate (100 mV to 10 V) 110 / sec
Execute simple command changes (CALL, OCOMP, etc.) 330 / sec
Readings to GPIB, ASCII 630 / sec
Readings to GPIB, DREAL 1000 / sec
Readings to GPIB, DINT 50,000 / sec
Readings to internal memory, DINT 50,000 / sec
Readings from internal memory to GPIB, DINT 50,000 / sec
Readings to GPIB, SINT 100,000 / sec
Readings to internal memory, SINT 100,000 / sec
Readings from internal memory to GPIB, SINT 100,000 / sec
Maximum internal trigger reading rate 100,000 / sec
Maximum external trigger reading rate 100,000 / sec
Standard Option 001
Readings Bytes Readings Bytes
Reading Storage (16 bit) 10,240 20 k +65,536 +128 k
Non-volatile, for subprograms
and / or state storage 14 k
Delay Time
Accuracy ±0.01% ± 5 ns
Maximum 6000 s
Resolution 10 ns
Jitter 50 ns pk-pk
Timer
Accuracy ±0.01% ±5 ns
Maximum 6000 s
Resolution 100 ns
Jitter <100 ps rms
1. Using HP 9000 Series
350.
2. SINT data is valid for
APER £10.8µs.