Appendix A Specifications 299
11 / General Specifications
Operating Environment
Temperature Range: 0°C to 55°C
Operating Location: Indoor Use Only
Operating Altitude: Up to 2,000 Meters
Pollution Rating: IEC 664 Degree 2
Operating Humidity Range
up to 95% RH at 40°C
Physical Characteristics
88.9 mm H x 425.5 mm W x 502.9 mm D
Net Weight: 12 kg (26.5 lbs)
Shipping Weight 14.8 kg (32.5 lbs)
Storage Temperature
–40°C to + 75°C
Warm-Up Time
4 Hours to published specifications
Power Requirements
100/120 V, 220/240 V ±10%
48–66Hz, 360–420Hz (auto sensed)
<30 W, <80 VA (peak)
Fused: 1.5 @ 115 V or 0.5 A @230 V
Cleaning Guidelines
To clean the instrument, use a clean
cloth slightly dampened with water.
Field Installation Kits Agilent Part Number
Option 001 Extended Reading Memory 03458-87901
Option 002 High Stability Reference 03458-80002
Extra Keyboard Overlays (5 each) 03458-84303
Available Documentation Agilent Part Number
Product Note 3458A-1: Optimizing Throughput and Reading Rate 5953-7058
Product Note 3458A-2: High Resolution Digitizing with the 3458A 5953-7059
Product Note 3458A-3: Electronic Calibration of the 3458A 5953-7060
Extra Manual Set 03458-90000
Warranty Period
One year
Input Terminals
Gold-plated Tellurium Copper
Input Limits
Input HI to LO: 300 Vac Max (CAT II)
IEEE-488 Interface
Complies with the following:
IEEE-488.1 Interface Standard
IEEE-728 Codes/Formats Standard
CIIL (Option 700)
Included with Agilent 3458A:
Test Lead Set (Agilent 34118A)
Power Cord
User’s Guide
Calibration Manual
Assembly Level Repair Manual
Quick Reference Guide