SLOPE
234 Chapter 6 Command Reference
sampling, or synchronous sampling. The parameters are:
Power-on type = ANA.
Default type = ANA.
Remarks Bandwidth limitations vary with the conversion technique selected. See the
Specifications in Appendix A for details.
Query Command. The SETACV? query command returns the present AC
measurement method. Refer to "Query Commands" near the front of this chapter
for more information.
Related Commands: ACBAND, ACDCV, ACV, FUNC, SSRC
Example 10 OUTPUT 722; "SETACV SYNC" !SPECIFIES SYNCHRONOUS SAMPLING (DC
COUPLED)
20 OUTPUT 722;"ACDCV" !SELECTS AC+DC VOLTAGE MEASUREMENTS
30 END
SLOPE
SLOPE is used in conjunction with the LEVEL command and specifies which
slope of the signal will be used by the level-detection circuitry.
Syntax SLOPE [slope]
slope
Selects the positive-going or negative-going slope of the input signal for use by
the level detection circuitry. The choices are:
Power-on slope = POS.
Default slope = POS.
Remarks Query Command. The SLOPE? query command returns the present slope.
type
Parameter
Numeric
Query
Equiv. Description
ANA 1 Analog RMS conversion
RNDM 2 Random sampling conversion
SYNC 3 Synchronous sampling conversion
slope
Parameter
Numeric
Query
Equiv. Description
NEG 0 Selects negative-going slope
POS 1 Selects positive-going slope