Chapter 4 Making Measurements 85
"High-Speed Mode" later in this chapter for more information. In the
following program, the PRESET NORM command sets the trigger event to
synchronous. Line 40 specifies 15 readings per synchronous trigger event.
Line 50 requests data from the multimeter. This satisfies the synchronous
trigger event and initiates the readings. Notice that line 50 requests data from
the multimeter 15 times. When multiple readings are specified and SYN is
used as the trigger or trigger arm event, the multimeter does not recognize
the multiple data requests as individual SYN events. That is, in this program
the SYN trigger event occurs once, not 15 times.
10 OPTION BASE 1 !COMPUTER ARRAY NUMBERING STARTS AT 1
20 DIM Rdgs (15) !DIMENSION ARRAY FOR 15 READINGS
30 OUTPUT 722;"PRESET NORM" !TARM AUTO, TRIG SYN, DCV AUTORANGE, MEM OFF
40 OUTPUT 722;"NRDGS 15,AUTO" !15 READINGS/TRIGGER, AUTO SAMPLE EVENT
50 ENTER 722;Rdgs(*) !GENERATE SYN EVENT, ENTER READINGS
60 PRINT Rdgs(*) !DISPLAY READINGS
70 END
The following program uses the synchronous event as the sample event. Line
60 requests data from the multimeter 15 times. When SYN is used as the
sample event, each request for data is recognized as a SYN event. That is, in
this program the SYN event occurs 15 times.
10 OPTION BASE 1 !COMPUTER ARRAY NUMBERING STARTS AT 1
20 DIM Rdgs(15) !DIMENSION ARRAY FOR 15 READINGS
30 OUTPUT 722;"PRESET NORM" !TARM AUTO, TRIG SYN, DCV AUTORANGE
40 OUTPUT 722;"NRDGS 15,SYN" !15 READINGS PER TRIGGER, SYN SAMPLE EVENT
50 OUTPUT 722;"TRIG AUTO" !AUTO TRIGGER EVENT
60 ENTER 722;Rdgs(*) !SYN EVENT, ENTER EACH READING
70 DISP Rdgs(*) !PRINT READINGS
80 END
Making Timed
Readings
When making multiple readings per trigger, you can use the TIMER sample
event to place a specified time interval between readings. This interval is the
amount of time from the beginning of one reading to the beginning of the
next reading. You specify the interval in seconds using the TIMER command.
(If the specified interval is less than the time required to make each reading,
the multimeter generates the TRlG TOO FAST error). The following
program specifies 8 readings per trigger with 1 second between readings (this
is shown in Figure 18).
10 OPTION BASE 1 !COMPUTER ARRAY NUMBERING STARTS AT 1
20 DIM Rdgs(8) !DIMENSION ARRAY FOR 8 READINGS
30 OUTPUT 722;"PRESET NORM" !TARM AUTO, TRIG SYN, DCV AUTORANGE
40 OUTPUT 722;"NRDGS 8, TIMER" !8 READINGS/TRIGGER, TIMER SAMPLE EVENT
50 OUTPUT 722;"TIMER 1" !1 SECOND TIMER INTERVAL
60 ENTER 722;Rdgs(*) !SYN EVENT,ENTER EACH READING
70 PRINT Rdgs(*) !PRINT READINGS
80 END
You can also use the SWEEP command to replace the NRDGS n,TIMER
command and the TlMER command. The SWEEP command's first
parameter specifies the interval between readings and its second parameter
specifies the number of readings. (The SWEEP and NRDGS commands are
interchangeable: the multimeter uses whichever was specified last in the
programming.) For example, the following program also takes 8 readings

Table of Contents

Other manuals for Agilent Technologies 3458A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3458A and is the answer not in the manual?

Agilent Technologies 3458A Specifications

General IconGeneral
BrandAgilent Technologies
Model3458A
CategoryMultimeter
LanguageEnglish

Related product manuals