EasyManua.ls Logo

Agilent Technologies 3458A

Agilent Technologies 3458A
372 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FUNC
184 Chapter 6 Command Reference
To select autorange, specify AUTO for max._input or default the parameter. In
the autorange mode, the multimeter samples the input signal before each reading
and selects the appropriate range.
The following tables show the max._input parameters and the ranges they select
for each measurement function.
Power-on max._input = AUTO.
Default max._input = AUTO.
For DCV:
For ACV or ACDCV:
For OHM or OHMF:
For DCI:
For ACI or ACDCI:
For DSAC or DSDC:
For SSAC or SSDC:
max._input
Parameter
Selects
Range
Full
Scale
–1 or AUTO
0 to .12
>.12 to 1.2
>1.2 to 12
>12 to 120
>120 to 1E3
Autorange
100mV
1V
10V
100V
1000V
120mV
1.2V
12V
120V
1050V
max._input
Parameter
Selects
Range
Full
Scale
–1 or AUTO
0 to .012
>.012 to .12
>.12 to 1.2
>1.2 to 12
>12 to 120
>120 to 1E3
Autorange
10mV
100mV
1V
10V
100V
1000V
12mV
120mV
1.2V
12V
120V
1050V
max._input
Parameter
Selects
Range
Full
Scale
–1 or AUTO
0 to 12
>12 to 120
>120 to 1.2E3
>1.2E3 to 1.2E4
>1.2E4 1.2E5
>1.2E5 to 1.2E6
>1.2E6 to 1.2E7
>1.2E7 1.2E8
>1.2E8 1.2E9
Autorange
109
1009
1k9
10k9
100k9
1M9
10M9
100M9
1G9
129
120k9
1.2k9
12k9
120k9
1.20M9
12M9
120M9
1.2G9
max._input
Parameter
Selects
Range
Full
Scale
–1 or AUTO
0 to .12E-6
>.12E-6 to 1.2E-6
>1.2E-6 to 12E-6
>12E-6 to 120E-6
>120E-6 to 1.2E-3
>1.2E-3 to 12E-3
>12E-3 to 120E-3
>120E-3 to 1.2
Autorange
.1µA
1µA
10µA
100µA
1mA
10mA
100mA
1A
.12µA
1.2µA
12µA
120µA
1.2mA
12mA
120mA
1.05A
max._input
Parameter
Selects
Range
Full
Scale
–1 or AUTO
0 to .120E-6
>120E-6 to 1.2E-3
>1.2E-3 to 12E-3
>12E-3 to 120E-3
>120E-3 to 1.2
Autorange
100µA
1mA
10mA
100mA
1A
120µA
1.2mA
12mA
120mA
1.05A
Full Scale
max._input
Parameter
Selects
Range
SINT
format
DINT
format
0 to .012
>.012 to .120
>.120 to 1.2
>1.2 to 12
>12 to 120
>120 to 1E3
10mV
100mV
1V
10V
100V
1000V
12mV
120mV
1.2V
12V
120V
1050V
50mV
500mV
5.0V
50V
500V
1050V
max._input
Parameter
Selects
Range
Full
Scale
0 to .012
>.012 to .120
>.120 to 1.2
>1.2 to 12
>12 to 120
>120 to 1E3
10mV
100mV
1V
10V
100V
1000V
12mV
120mV
1.2V
12V
120V
1050V

Table of Contents