PRESET
218 Chapter 6 Command Reference
AZERO ON MFORMAT SREAL
BEEP ON MMATH OFF
DCV AUTO NDIG 6
DELAY -1 NPLC 1
DISP ON NRDGS 1,AUTO
FIXEDZ OFF OCOMP OFF
FSOURCE ACV OFORMAT ASCII
INBUF OFF TARM AUTO
LOCK OFF TIMER 1
MATH OFF TRIG SYN
All math registers set to 0 except:
DEGREE = 20
PERC = 1
REF = 1
RES = 50
SCALE = 1
FAST
PRESET FAST configures the multimeter for fast readings, fast transfer to
memory, and fast transfer from memory to GPIB. (Refer to "Increasing the
Reading Rate" in Chapter 4 for more information on fast measurements.)
Executing PRESET FAST executes the commands shown under PRESET NORM
with the following exceptions:
DCV 10
AZERO OFF
DISP OFF
MFORMAT DINT
OFORMAT DINT
TARM SYN
TRIG AUTO
DIG
PRESET DIG configures the multimeter for DCV digitizing (DCV digitizing is
discussed in Chapter 5). Executing PRESET DIG executes the commands shown
under PRESET NORM with the following exceptions:
DCV 10
AZERO OFF
DELAY 0
DISP OFF
TARM HOLD
TRIG LEVEL
LEVEL 0,AC
NRDGS 256,TIMER
TIMER 20E-6
APER 3E-6
MFORMAT SINT