78 Chapter 3 Configuring for Measurements
enabled still respond to their corresponding conditions. They do not,
however, set bit 6 or assert SRQ. The following program is an example of
interrupts using HP Series 200/300 BASIC.
10 !HI/LO LIMIT EXCEEDED,ERROR, POWER CYCLED INTERRUPT
20 OUTPUT 722;"PRESET NORM"
30 OUTPUT 722; "CSB"
40 ON INTR 7 GOTO 90
50 ENABLE INTR 7;2
60 OUTPUT 722;"RQS 42;MATH PFAIL;SMATH MIN -5;SMATH MAX
5"
70 OUTPUT 722;"TRIG AUTO"
80 GOTO 80
90 OUTPUT 722; "STB?"
100 ENTER 722;A
110 IF BINAND (A,2) THEN PRINT "HI/LO LIMIT EXCEEDED"
120 IF BINAND (A,8) THEN PRINT "POWER WAS CYCLED"
130 IF BINAND (A,32) THEN PRINT "ERROR OCCURRED"
140 END
Line 20 presets the multimeter, which suspends triggering. Line 30 clears the
status register. Line 40 instructs the controller to go to line 90 should an
interrupt occur. Line 50 enables SRQ interrupts on the GPIB interface. Line
60 enables the hi/lo limit, power-on, and error bits to assert SRQ. Line 60
also enables the real-time pass/fail math operation with the values of -5 for
the low limit and +5 for the hi limit. Line 70 enables automatic triggering.
Line 80 causes the controller to wait for an interrupt. Lines 90 through 130
read the status register and print which condition(s) caused the interrupt.

Table of Contents

Other manuals for Agilent Technologies 3458A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3458A and is the answer not in the manual?

Agilent Technologies 3458A Specifications

General IconGeneral
BrandAgilent Technologies
Model3458A
CategoryMultimeter
LanguageEnglish

Related product manuals