165
Specifications
Definition of Terms
C.1 Definition of Terms 
Attenuation accuracy
The difference between the displayed loss and 
→excess loss.
Conditions: Attenuation adjustment prior to measurement. That is, 
adjustment of the measured attenuation at the highest setting so that 
it equals the attenuation setting, for example by adjusting the 
wavelength setting.
Measurement: with laser source or LED and optical power meter.
Attenuation range
The range of displayed attenuations.
Excess loss
The difference between actual loss (at an arbitrary attenuation 
setting) and [rightarrow]insertion loss (at 0 dB setting).
Insertion loss
The change of power levels after inserting the attenuator between 
two connectorized patchcords, with the attenuation set to 0 dB.
Conditions: Arbitrary wavelength setting, temperature within 
operating temperature range, jumper cables with high quality 
connectors.
Measurement: with laser source or LED and optical power.
Polarization dependent loss
The dependence of the attenuation on the input polarization state, 
expressed as the difference between the highest and the lowest 
displayed attenuation, in dB.
Conditions: Fabry-Perot type laser source with variable 
polarization state and polarization-independent power, generation 
of all polarization states (covering the entire Poincar sphere), 
jumper cables with high-quality connectors.