40 
Using the Attenuator
Setting Up the Attenuation
1. press CAL, and
2. edit the factor using the Modify keys (see “Using the Modify 
Keys” on page 29).
Resetting the Calibration Factor 
To reset the calibration factor, press and hold C
AL until the value 
resets to zero (this takes approximately two seconds). The 
calibration factor resets to zero.
Transferring to the Calibration Factor 
You can transfer the attenuation factor shown on the display into 
the calibration factor, so that the attenuation factor is reset to zero.
You would use this, for example, after you have set the power 
through the attenuator at a specific level. When you have reset the 
attenuation factor, you can edit it to get a relative attenuation.
The filter attenuation stays constant when you transfer to the 
calibration factor. This means that the new calibration factor is 
calculated from the attenuation factor and the old calibration factor 
according to the formula below (from equation (1)):
Cal
NEW
(dB) = -Att
filter
(dB) = Cal
OLD
(dB) - Att
OLD
(dB)
To transfer to the calibration factor, press D
ISP→CAL.
Entering the Wavelength 
The attenuation at any point on the filter is wavelength dependent. 
This dependence is measured and stored in the instrument, and is 
used, with the value for the wavelength entered by the user, to 
compensate for the dependence. This is the wavelength calibration 
data.
NOTE There are two ways of using the wavelength calibration data,
• to reposition the filter so that the attenuation stays constant, or
• to change the attenuation factor on the display to show the 
wavelength dependence. You use this to set the wavelength for an 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com