EasyManua.ls Logo

Agilent Technologies 8156A - Replacing the Battery; Replacing the Fuse

Agilent Technologies 8156A
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
146
Installation
AC Line Power Supply Requirements
Replacing the Battery
This instrument contains a lithium battery. Replacing thebattery
should be carried out only by a qualified electrician or by Agilent
Technologies service personnel.
There is a danger of explosion if the battery is incorrectly replaced.
Replace only with the same or an equivalent type (Agilent part
number 1420-0394). Discard used batteries according to local
regulations.
Replacing the Fuse
There is one fuse in this instrument. This is a T1A/250V (time-lag)
(Agilent Part No. 2110-0007). The fuse holder is at the rear of the
instrument, beside the line power connector. To replace the fuse,
1. Release the fuse holder: use the blade of a flat-headed
screwdriver to depress the catch at the side of the holder and then
pull the holder out a little.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Related product manuals