Making an Attenuation Sweep
The Manual Sweep
Resetting the Parameters
To reset any of the sweep parameters, press and hold SWP until the
value resets (this takes approximately two seconds).
START and STOP reset so that the filter attenuation (inside the
instrument) is zero, that is
Start = Cal
or
Stop = Cal
See Entering a Calibration Factor on page 39 for information
about setting the calibration factor, Cal.
STEP resets to zero.
Executing the Manual Sweep
If you have just set up your sweep, then you only need to press
E
XEC to run the application.
If you have already set up the sweep, and are currently operating
the instrument as an attenuator,
1. Press S
WP, and then,
2. Press E
XEC.
Figure 3-6 Running the Manual Sweep
If there is something wrong with a parameter (if STEP is zero, for
example), this parameter is shown on the display for editing. Edit
the parameter, and press E
XEC again.