EasyManua.ls Logo

Agilent Technologies 8156A - Figure 2-6 Hardware Configuration for Attenuation Example - B

Agilent Technologies 8156A
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
43
Using the Attenuator
Example, Setting the Calibration
NOTE Under normal circumstances you should leave the instruments to
warmup. (The multimeter needs around 20 minutes to warmup. The
attenuator needs around 45 minutes with the shutter open to warmup.)
Warming up is necessary for accuracy of the sensor, and the output
power of the source.
d. Connect a patchcord from the source to the input of the
sensor.
2. Measure the insertion loss of the Hardware setup:
a. On the multimeter:
i. Set the wavelength for the sensor to that of the source.
ii. Activate the source, by pressing the gray button on its
front panel.
iii. Start the loss application (press M
ODE and then LOSS,
and E
XEC).
b. Reconfigure the hardware to include the attenuator:
i. Disconnect the source from the sensor, and connect it to
the input of the attenuator.
Figure 2-6 Hardware Configuration for Attenuation Example - B
ii. Connect a patchcord from the output of the attenuator to
the sensor.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Related product manuals