107
Remote Commands
INPut Commands
OUTPUT 728;":INP:ATT?" 
ENTER 728;A$
:INPut:LCMode 
Syntax     :INPut:LCMode <wsp> OFF|ON|0|1
Description     This command sets the function of the 
wavelength calibration. That is, whether the 
wavelength calibration data is to be used to 
reposition the filter to keep the attenuation 
factor constant, or to alter the attenuation factor 
with the filter kept in a fixed position.
Switch the mode on (using OFF or 0) to keep 
the attenuation value fixed, and alter the filter 
position. Switch the mode off (using ON or 1) 
to keep the filter position fixed, and alter the 
attenuation factor.
:INPut:LCMode? 
Syntax     :INPut:LCMode?
Description     The query returns the current function of the 
wavelength calibration.
0 indicates that the instrument is keeping the 
attenuation value fixed, and altering the filter 
position. 1 indicates the instrument is keeping 
the filter position fixed, and altering the 
attenuation factor.
Example     OUTPUT 728;":INP:LCM ON"
OUTPUT 728;":INP:LCM?" 
ENTER 728;A$
:INPut:OFFSet 
Syntax     :INPut:OFFSet <wsp> 
<value>[DB]|MIN|DEF|MAX