EasyManuals Logo

Agilent Technologies 8156A User Manual

Agilent Technologies 8156A
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #276 background imageLoading...
Page #276 background image
275
Error Messages
Display Messages
F.1 Display Messages
FAILnnnn
indicates that the self test has failed. The number nnnn is a four
digit hexadecimal number that indicates which part of the self test
has failed.
So FAIL0010 would mean that the DSP (Digital Signal Processor)
Communications had failed, FAIL0012 would mean that the DSP
Communications had failed, and so had the Battery RAM.
Bits Mnemonics
Hexadecimal
Value
8 Counter 0100
16
7 Analog to Digital Convertor 0080
16
6 General DSP Hardware 0040
16
5 DSP Timeout 0020
16
4 DSP Communications 0010
16
3 Calibration Data 0008
16
2 Keypad 0004
16
1 Battery RAM 0002
16
0 Calibration Data Checksum 0001
16