Remote Commands
The Common Commands
(High performance, high return loss version),
the string returned is High Performance,
0,
High Return Loss.
Example OUTPUT 728;"*OPT?"
ENTER 728;A$
*RCL
Syntax *RCL <wsp> <location>
0
location 9
Definition An instrument setting from the internal RAM is
made the actual instrument setting (this does
not include GPIB address or parser, the
attenuation resolution or the power on setting).
You recall user settings from locations 1-9. See
*SAV on page 100. Location 0 contains the
default setting, which is the same as that
obtained by *RST.
Example OUTPUT 728;"*RCL 3"
*RST
Syntax *RST
Definition The reset setting (default setting) stored in
ROM is made the actual setting.
Instrument state: the instrument is placed in the
idle state awaiting a command.
The following are not changed:
GPIB (interface) state
Instrument interface address
Output queue
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8156A and is the answer not in the manual?

Agilent Technologies 8156A Specifications

General IconGeneral
BrandAgilent Technologies
Model8156A
CategoryTest Equipment
LanguageEnglish

Related product manuals