282 
Error Messages
GPIB Messages
-314 Save/recall memory lost 
The nonvolatile data saved by the *SAV command has been lost.
-315 Configuration memory lost 
The nonvolatile configuration data saved by the instrument has 
been lost.
-330 Self-test failed 
Further information about the self-test failure is available by using 
*TST?.
-350 Queue overflow 
The error queue has overflown. This error is written to the last 
position in the queue, no further errors are recorded.
Query Errors 
These are error messages in the range -400 to -499. They indicate 
that an error has been detected by the output queue control.
An device-specific error is signaled by the query error bit (bit 2) in 
the event status register.
-300 Query error 
This indicates that a query error has occurred. No more specific 
information is available.
-410 Query INTERRUPTED 
A condition occurred which interrupted the transmission of the 
response to a query (for example, a query followed by a DAB or a 
GET before the response was completely sent).
-420 Query UNTERMINATED 
A condition occurred that interrupted the reception of a query (for 
example, the instrument was addressed to talk and an incomplete 
program message was received).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com