Cleaning Information
How to clean connector adapters
1. Moisten a new cotton-swab with isopropyl alcohol.
2. Clean the connector by rubbing the cotton-swab over the surface
using a small circular movement.
3. Take a new, dry soft-tissue and remove the alcohol, dissolved
sediment and dust, by rubbing gently over the surface using a
small circular movement.
4. Blow away any remaining lint with compressed air.
An Alternative Procedure
A better, more gentle, but more expensive cleaning procedure is to
use an ultrasonic bath with isopropyl alcohol.
1. Hold the tip of the connector in the bath for at least three
minutes.
2. Take a new, dry soft-tissue and remove the alcohol, dissolved
sediment and dust, by rubbing gently over the surface using a
small circular movement.
3. Blow away any remaining lint with compressed air.
E.8 How to clean connector adapters
CAUTION Some adapters have an anti-reflection coating on the back to reduce
back reflection. This coating is extremely sensitive to solvents and
mechanical abrasion. Extra care is needed when cleaning these
adapters.
Preferred Procedure
Use the following procedure on most occasions.