102 Agilent 1290 Infinity Binary Pump User Manual
8 Test Functions and Calibrations
Introduction
Introduction
All tests are described based on the Agilent Lab Advisor Software B.01.04.
Other user interfaces may not provide any test or just a few.
For details on the use of the interface refer to the interface documentation.
Tab le 8 Interfaces and available test functions
Interface Comment Available Function
Agilent Instrument Utilities Maintenance tests available System pressure test
Agilent Lab Advisor All tests are available System pressure test
Pump head leak test
Agilent ChemStation No tests available
Adding of pressure to
chromatographic signals
possible
Pressure
Pressure ripple
Temperature main board
Agilent Instant Pilot Some tests are available System pressure test
Monitoring of values
Pressure
Pressure ripple
Flow (in case of
operating pressure)

Table of Contents