160 Agilent 1290 Infinity Binary Pump User Manual
9 Maintenance
Replacing the Main Power Fuses
Replacing the Main Power Fuses
When If the main power LED is off while the main power button is pressed (see “Status indicators” on
page 71).
Tools required Description
Flat head screwdriver
Parts required # p/n Description
1 2110-1004 Fuse 10 A t
Preparations Switch off the instrument and unplug the main power cable.
WARNING
Fire hazard
Using wrong fuses can result in fire hazard.
For continued protection against fire hazard, replace line fuses only with the same
type and ratings.
Only use the fuses specified for this instrument. The use of other fuses or materials
is prohibited.
NOTE
There are more fuses inside the instrument.
If replacing the main power fuse does not resolve the issue, please contact your Agilent
service representative.