Agilent 1290 Infinity Binary Pump User Manual 85
Error Information
7
Pump Error Messages
Pump Error Messages
These errors are pump specific.
Pressure of binary pump above upper limit
Error ID: 22014
The pressure has exceeded the upper pressure limit.
Parameter: Measured pressure
Probable cause Suggested actions
1 Blockage in flow path after the pressure
sensor.
Check for blockages in the LC system, e.g.
purge valve, Jet Weaver, degraded column,
column frits, needle, needle seat, capillaries
etc.
Check for particles in the solvent.
2 Inappropriate settings (pressure limit, flow
rate).
Decrease flow rate.
Increase pressure limit.