Pump Error Messages
Binary pump shutdown during analysis
Error ID: 22065
The binary pump has been shut down by the control software or control
module during an analysis.
• Parameter: 0 for off, 1 for standby.
Reading the pump encoder tag failed
Error ID: 22402
Reading the pump encoder tag has failed.
• Parameter: 1 – 4 referring to pump drive
Writing the pump encoder tag failed
Error ID: 22405
Writing the pump encoder tag has failed.
• Parameter: 1 – 4 referring to pump drive
Probable cause Suggested actions
1 Pump has been shut down.
Restart pump.
Probable cause Suggested actions
1 Defect connection between encoder and
main board.
Please contact your Agilent service
representative.
2 Missing or defect tag Defect connection
between tag and encoder.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defect connection between encoder and
main board.
Please contact your Agilent service
representative.
2 Defect tag Defect connection between tag
and encoder.
Please contact your Agilent service
representative.