Agilent 1290 Infinity Binary Pump User Manual 89
Error Information
7
Pump Error Messages
Binary pump shutdown during analysis
Error ID: 22065
The binary pump has been shut down by the control software or control
module during an analysis.
Parameter: 0 for off, 1 for standby.
Reading the pump encoder tag failed
Error ID: 22402
Reading the pump encoder tag has failed.
Parameter: 1 – 4 referring to pump drive
Writing the pump encoder tag failed
Error ID: 22405
Writing the pump encoder tag has failed.
Parameter: 1 – 4 referring to pump drive
Probable cause Suggested actions
1 Pump has been shut down.
Restart pump.
Probable cause Suggested actions
1 Defect connection between encoder and
main board.
Please contact your Agilent service
representative.
2 Missing or defect tag Defect connection
between tag and encoder.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defect connection between encoder and
main board.
Please contact your Agilent service
representative.
2 Defect tag Defect connection between tag
and encoder.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals