234 Agilent 1290 Infinity Binary Pump User Manual
13 LAN Configuration
Dynamic Host Configuration Protocol (DHCP)
Setup (DHCP)
1 Note the MAC address of the LAN interface (provided with G1369C LAN
Interface Card or Main Board). This MAC address is on a label on the card
or at the rear of the main board, e.g. 0030d3177321.
On the Instant Pilot the MAC address can be found under Details in the LAN
section.
Figure 42 LAN Setting on Instant Pilot
2 Set the Configuration Switch to DHCP either on the G1369C LAN Interface
Card or the main board of above mentioned modules.
Software required The modules in the stack must have at least firmware from set A.06.34 and the above mentioned
modules B.06.40 or above (must from the same firmware set).
Tab le 18 G1369C LAN Interface Card (configuration switch on the card)
SW 4 SW 5 SW 6 SW 7 SW 8 Initialization Mode
ON OFF OFF OFF OFF DHCP