Algae Growth in HPLC Systems
Algae Growth in HPLC Systems
The presence of algae in HPLC systems can cause a variety of problems that
may be incorrectly diagnosed as instrument or application problems. Algae
grow in aqueous media, preferably in a pH range of 4-8. Their growth is
accelerated by buffers, for example phosphate or acetate. Since algae grow
through photosynthesis, light will also stimulate their growth. Even in distilled
water small-sized algae grow after some time.
Instrumental Problems Associated With Algae
Algae deposit and grow everywhere within the HPLC system causing:
Blocked solvent filters or deposits on inlet or outlet valves resulting in
unstable flow, composition or gradient problems or a complete failure of
the pump.
Small pore high pressure solvent filters, usually placed before the injector
to plug resulting in high system pressure.
PTFE frits blockage leading to increased system pressure.
Column filters to plug giving high system pressure.
Flow cell windows of detectors to become dirty resulting in higher noise
levels (since the detector is the last module in the flow path, this problem is
less common).
How to Prevent and-or Reduce the Algae Problem
Always use freshly prepared solvents, especially use demineralized water
which was filtered through about 0.2 µm filters.
Never leave mobile phase in the instrument for several days without flow.
Always discard old mobile phase.
Use the amber solvent bottle (Solvent bottle, amber (9301-1450)) supplied
with the instrument for your aqueous mobile phase.
If possible add a few mg/l sodium azide or a few percent organic solvent to
the aqueous mobile phase.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals