92 Agilent 1290 Infinity Binary Pump User Manual
7 Error Information
Pump Error Messages
Overcurrent of solvent selection valve (SSV)
Error ID: 22412
Overcurrent of solvent selection valve (SSV).
Parameter: None
Deliver underrun
Error ID: 22413
Internal error.
Parameter: None
Defect connection between main board and pump drive encoder
Error ID: 22414
Defect connection between main board and pump drive encoder.
Parameter: 1 – 4 referring to pump drive
Probable cause Suggested actions
1 Valve defect.
Replace the solvent selection valve.
Probable cause Suggested actions
1 Internal error.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defect connection between main board and
pump drive encoder.
Please contact your Agilent service
representative.
2 Defect encoder.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals