Agilent 1290 Infinity Binary Pump User Manual 99
Error Information
7
Pump Error Messages
Seal wash pump was missing when tried to turn on
Error ID: 22499
The seal wash pump has not been detected (while being configured or detected
before)
Probable cause Suggested actions
1 Defect cable connection to seal wash pump.
Check cable connection.
2 Defect seal wash pump motor.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.