12 Agilent 1290 Infinity Binary Pump User Manual
1 Introduction
Pump Principle
Pump Principle
The Binary Pump is based on a two-channel, dual-piston in-series design
which comprises all essential functions that a solvent delivery system has to
fulfill. Metering of solvent and delivery to the high-pressure side are
performed by two pump assemblies which can generate pressure up to
1200 bar.
Each channel comprises a pump assembly including independent pump drive
for each piston, pump head, inlet valve, outlet valve, solvent heat exchanger
and an outlet filter. The two channels are fed into a low-volume mixing groove
in an automatic purge valve and a Jet Weaver mixer, with 35 or 100 µl volume
can be added downstream for optimum mixing performance.
A system pressure sensor, for monitoring the pump pressure, is attached to
the purge valve, normally connected in the B-channel of the pump, before the
mixing groove, in order to minimize delay volumes.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals