228 Agilent 1290 Infinity Binary Pump User Manual
13 LAN Configuration
Initialization mode selection
Initialization mode selection
The following initialization (init) modes are selectable:
Bootp
When the initialization mode Bootp is selected, the module tries to download
the parameters from a Bootp Server. The parameters obtained become the
active parameters immediately. They are not stored to the non-volatile
memory of the module. Therefore, the parameters are lost with the next power
cycle of the module.
Figure 37 Bootp (Principle)
Tab le 16 Initialization Mode Switches
SW 6SW 7SW 8Init Mode
OFF OFF OFF Bootp
OFF OFF ON Bootp & Store
OFF ON OFF Using Stored
OFF ON ON Using Default
ON OFF OFF
DHCP
1
1
Requires firmware B.06.40 or above. Modules without LAN on board, see G1369C LAN Interface Card