88 Agilent 1290 Infinity Binary Pump User Manual
7 Error Information
Pump Error Messages
Waste counter limit exceeded
Error ID: 22056
The counter for the waste volume has exceeded the limit, which has been set
in the user interface.
Parameter: None
Flow rate limit exceeded
Error ID: 22064
The flow rate of the binary pump has exceeded the limit, while the pump runs
in pressure controlled mode, e.g. during a pressure test.
Parameter: None
Probable cause Suggested actions
1 The waste container is full.
Empty waste container.
2 Inappropriate setting for waste counter.
Reset waste counter.
Adjust waste counter limit.
Probable cause Suggested actions
1 Leak
Check for leaks in the pump and flow path.
2 Bottle empty.
Fill solvent bottle.
3 Shutoff valve closed.
Open shutoff valve.
4 Drift of pressure sensor (unlikely for short
tests taking some minutes).
Replace pressure sensor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals