EasyManuals Logo

Agilent Technologies Infinity Binary Pump VL 1290 User Manual

Agilent Technologies Infinity Binary Pump VL 1290
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
90 Agilent 1290 Infinity Binary Pump User Manual
7 Error Information
Pump Error Messages
Pump drive blocked or encoder failed
Error ID: 22406
Pump drive blocked or encoder failed.
Parameter: None
Drive current too low
Error ID: 22407
The current consumption of the pump drive is too low.
Parameter: 1 – 4 referring to pump drive
Drive Encoder failed
Error ID: 22408
Drive encoder failed during pump drive calibration.
Probable cause Suggested actions
1 Blockage of the pump drive Drive encoder
failed.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Drive motor defect.
Please contact your Agilent service
representative.
2 Wrong/missing connection of pump drive to
main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Internal error.
Contact Agilent support.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals