Agilent 1290 Infinity Binary Pump User Manual 87
Error Information
7
Pump Error Messages
Degasser's pressure limit violation
Error ID: 22032
Pressure too far above the limit.
Solvent counter exceeded limit
Error ID: 22055
The counter for the solvent volume has exceeded the limit, which has been set
in the user interface.
Parameter:
Without Solvent Selection Valve:
0 for channel A, 1 for channel B
With Solvent Selection Valve:
2 for channel A1, 3 for channel B1, 4 for channel A2, 5 for channel B2
Probable cause Suggested actions
1 Leak in degasser chamber or degasser
tubing.
Please contact your Agilent service
representative.
2 Defect vacuum pump.
Please contact your Agilent service
representative.
3 Degasser chamber empty or connected to
air.
Block unused degasser channels.
Probable cause Suggested actions
1 No solvent present.
Refill solvent bottle.
2 Inappropriate setting.
Check solvent counter setting in user interface.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals