96 Agilent 1290 Infinity Binary Pump User Manual
7 Error Information
Pump Error Messages
Current of primary pump drive too high
Error ID: 22433
The current of the primary pump drive is too high.
Parameter: 1 or 4 referring to pump drive.
Current of secondary pump drive too high
Error ID: 22434
The current of the secondary pump drive is too high.
Parameter: 2 or 3 referring to pump drive
Probable cause Suggested actions
1 Blockage of flow path between primary
pump head and pressure sensor, e.g. of the
heat exchanger.
Check for blockages in flow path.
Please contact your Agilent service
representative.
2 Primary pump drive is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Blockage of flow path between secondary
pump head and pressure sensor, e.g. of the
heat exchanger.
Check for blockages in the flow path.
Please contact your Agilent service
representative.
2 Secondary pump drive is defect.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals