Site Requirements and Specifications
2
Physical Specifications
Physical Specifications
Tab le 1 Physical Specifications
Type Specification Comments
Weight 21.8 kg (48 lbs)
Dimensions
(height × width × depth)
240 x 345 x 435 mm
(9.3 x 13.5 x 17 inches)
Line voltage 100 – 240 VAC, ± 10 % Wide-ranging
capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 350 VA / 270 W / 922 BTU Maximum
Ambient operating
temperature
4–55 °C (41–131 °F)
Ambient non-operating
temperature
-40 – 70 °C (-4 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 2000 m (6562 ft)
Non-operating altitude Up to 4600 m (15091 ft) For storing the module
Safety standards:
IEC, CSA, UL
Installation category II, Pollution degree 2 For indoor use only.