Agilent 1290 Infinity Binary Pump User Manual 91
Error Information
7
Pump Error Messages
Drive current too high
Error ID: 22409
The current consumption of the pump drive is too high.
Parameter: 1 – 4 referring to pump drive
Drive timeout
Error ID: 22410
Drive is blocked mechanically, fails during initialization.
Parameter: 1 – 4 referring to pump drive
Overcurrent of pump drive
Error ID: 22411
The current consumption of the pump drive is too high.
Parameter: 1 – 4 referring to pump drive
Probable cause Suggested actions
1 Blockage of system before pressure sensor.
Check for blockage of e.g. outlet valve filter frit,
purge valve, heat exchanger.
2 Drive motor defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Blockage of pump drive Drive motor defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Blockage of system before pressure sensor.
Check for blockage of e.g. outlet valve filter frit,
purge valve, heat exchanger.
2 Drive motor defect.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals