Agilent 1290 Infinity Binary Pump User Manual 97
Error Information
7
Pump Error Messages
Unknown purge valve type
Error ID: 22435
The type information of the purge valve is invalid.
Parameter: None
Pump drive encoder error
Error ID: 22437
The pump drive encoder has generated no signal.
Parameter: 1 – 4 referring to pump drive
Pump drive error
Error ID: 22438, 22439
The pump drive failed during calibration.
Parameter: 1 – 4 referring to pump drive
Probable cause Suggested actions
1 Wrong valve head installed.
Check or replace purge valve head.
2 Valve head has invalid RFID tag content.
Check or replace purge valve head.
Probable cause Suggested actions
1 Pump drive encoder is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Pump drive motor defect or mechanically
blocked.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals