98 Agilent 1290 Infinity Binary Pump User Manual
7 Error Information
Pump Error Messages
Pump drive stroke blocked
Error ID: 22441
The pump drive movement is blocked.
Parameter: 1 – 4 referring to pump drive
Pump drive stop not found
Error ID: 22442
The pump drive stop has not been found.
Parameter: 1 – 4 referring to pump drive
Pressure sensor calibration wrong or missing
Error ID: 22443
Pressure sensor calibration wrong or missing.
Parameter: None
Probable cause Suggested actions
1 Pump drive motor is mechanically blocked.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Pump drive spindle is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Pressure sensor calibration wrong or
missing.
Replace pressure sensor.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals