218 Agilent 1290 Infinity Binary Pump User Manual
12 Hardware Information
Setting the 8-bit Configuration Switch
Special Settings
The special settings are required for specific actions (normally in a service
case).
Boot-Resident
Firmware update procedures may require this mode in case of firmware
loading errors (main firmware part).
If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident mode. It is not operable as a
module. It only uses basic functions of the operating system for example, for
communication. In this mode the main firmware can be loaded (using update
utilities).
NOTE
The tables include both settings for modules – with on-board LAN and without on-board
LAN. They are identified as LAN and no LAN.
Tab le 13 Boot Resident Settings (On-board LAN)
Mode Select SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
TEST/BOOT11100000

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals