EasyManua.ls Logo

Agilent Technologies Infinity Binary Pump VL 1290

Agilent Technologies Infinity Binary Pump VL 1290
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1290 Infinity Binary Pump User Manual 219
Hardware Information
12
Setting the 8-bit Configuration Switch
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
CAUTION
Loss of data
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are calibration settings, diagnosis and repair log books which will not be
erased.
Save your methods and data before executing a forced cold start.
Tab le 14 Forced Cold Start Settings (On-board LAN)
Mode Select SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
TEST/BOOT11000001

Table of Contents

Related product manuals