Agilent 1290 Infinity Binary Pump User Manual 219
Hardware Information
12
Setting the 8-bit Configuration Switch
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
CAUTION
Loss of data
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are calibration settings, diagnosis and repair log books which will not be
erased.
Save your methods and data before executing a forced cold start.
Tab le 14 Forced Cold Start Settings (On-board LAN)
Mode Select SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
TEST/BOOT11000001