Agilent 1290 Infinity Binary Pump User Manual 127
Maintenance
9
Changing configuration or replacing the Jet Weaver
Changing configuration or replacing the Jet Weaver
When For optimizing the pump configuration to mixing performance or low delay volumes/fast gradients,
see chapter Optimizing Performance.
Tools required p/n Description
8710-0510 ¼ inch wrench
3mm hex key
Parts required # p/n Description
1 G4220-60006 Jet Weaver 35 µL/100 µL
1 G4220-60012 Jet Weaver 380 µL (OPTIONAL)
1 G4220-87000 Capillary SST Valve to Jet Weaver 300 mm x 0.17 mm I.D.