104 Agilent 1290 Infinity Binary Pump User Manual
8 Test Functions and Calibrations
System Pressure Test
Figure 12 System Pressure Test – dynamic pressure input
System Pressure Test Evaluation
Test Failed
Probable cause Suggested actions
1 Damaged blank nut (poorly shaped from
over tightening)
Before investigating any other possible sources
of failure make sure that the blank nut you are
using is in a good condition and properly
tightened.
2 Pump leakages
Perform the Pump Head Leak test.
3 Loose or leaky fittings
Tighten the fittings or replace capillaries.
4 Autosampler leakages
Perform the Autosampler Leak test.
5 Themostatted Column Compartment valve
leakages
Replace the TCC valve rotor seal.
Notice the difference between error in the test and a failed result! An error is caused by an
abnormal termination during the operation of the test, whereas a failed result indicates that
the test result were not within the specified limits.