56 Agilent 1290 Infinity Binary Pump User Manual
4 Using the Pump
Solvent Information
Solvent Information
Introduction
Observe the following recommendations on the use of solvents.
Follow recommendations for avoiding the growth of algae, see “Algae
Growth in HPLC Systems” on page 43.
Small particles can permanently block capillaries and valves. Therefore,
always filter solvents through 0.4 µm filters.
Avoid or minimize the use of solvents that may corrode parts in the flow
path. Consider specifications for the pH range given for different materials
like flow cells, valve materials etc. and recommendations in subsequent
sections.
Materials in Flow Path
Following materials are used in the flow path of this module:
Part Materials
Degasser chamber TFE/PDD Copolymer, FEP, PEEK, PPS
Microfluidic structures
1
1
Jet Weaver, Heat Exchanger
SST
SSV PEEK, FEP, PFA, Al
2
O
3
-based ceramic, ruby, sapphire, SST
Passive inlet valve SST, gold, ruby, ZrO
2
-based ceramic, tantalum
Outlet valve SST, gold, ruby, ZrO
2
-based ceramic, tantalum
Pump head SST
Pistons ZrO
2
-based ceramic
Piston/wash seals UHMW-PE, SST
Pressure sensor SST
Automatic purge valve Polyimide, SST, DLC

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals