146 Agilent 1290 Infinity Binary Pump User Manual
9 Maintenance
Assembling the Pump Head
Assembling the Pump Head
See chapter "Parts" for details.
When Before installing the pump head.
Tools required p/n Description
Pump head alignment tool
G4220-20012 Torque wrench 2 – 25 Nm
G4220-20013 4 mm hex bit
G4220-20014 2.5 mm Hex Bit
G4220-20015 Adapter ¼ in square to hex
01018-23702 Insert tool
Parts required p/n Description
0905-1420 PE seals (pack of 2)
0905-1718 Wash Seal PE
5062-2484 Gasket, seal wash (pack of 6)
CAUTION
Limited life time of the pump head
Inserting the backup seal wrongly may limit the life time of the pump head.
Please note the correct orientation of the backup seal.
CAUTION
Damage of the pump piston
The pump piston is very sensitive to shearing forces from the side.
Use the alignment piston of the pump head alignment tool for the alignment
procedure described below.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals