Agilent 1290 Infinity Binary Pump User Manual 121
Maintenance
9
Replacing the Inlet valve
3
With a 14 mm wrench, unscrew the inlet valve and
remove it.
4 Install inlet valve and tighten it at 10 Nm with a torque
wrench (14 mm).
5 Attach the inlet tubing at the inlet valve. 6 Open the shut off valves and purge the system to remove
air.
10 Nm