EasyManua.ls Logo

Agilent Technologies Infinity Binary Pump VL 1290

Agilent Technologies Infinity Binary Pump VL 1290
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1290 Infinity Binary Pump User Manual 121
Maintenance
9
Replacing the Inlet valve
3
With a 14 mm wrench, unscrew the inlet valve and
remove it.
4 Install inlet valve and tighten it at 10 Nm with a torque
wrench (14 mm).
5 Attach the inlet tubing at the inlet valve. 6 Open the shut off valves and purge the system to remove
air.
10 Nm

Table of Contents

Related product manuals