Agilent 1290 Infinity Binary Pump User Manual 47
Using the Pump
4
Setting up the Pump with the Instrument Control Interface
Instrument Signals
The following pump signals are displayed:
Context Menu
The context menu of the dashboard panel contains the following commands:
Flow The current solvent flow rate (in mL/min).
Pressure The current pump ressure (in bar, psi or MPa, see āInstrument
Configurationā on page 45).
Pressure Limit The current maximum pressure limit.
Composition A:B The current solvent composition. When a solvent selection valve is fitted,
the channels are shown in the graphic.
Control Displays the pump's Control dialog box.
Method Displays the pump's Method Setup dialog box.
Set Error Method Sets the method that is loaded if an error occurs to the
method that is currently available in the hardware.
Identify Device Causes the LED on the front of the module to blink for a
few seconds.
Switch Pump On/Off Toggles the status of the pump, on or off.
Bottle Fillings Displays the Bottle Fillings dialog box.
Purge On/Off Allows you to control the purging of the system.
Prime On/Off Allows you to prime the pump heads for initially drawing
solvent.
Conditioning On/Off Allows you to switch pump conditioning on and off. The
conditioning function is useful for removing small air
bubbles inside the pump flow path.